The Automated Blood Analyzer AM-BAA10 by Abimed is engineered for precise and efficient blood analysis, featuring compact dimensions of 410 mm × 300 mm × 400 mm. It functions in two modes: whole blood, with a sample volume of up to 9 microliters, and prediluted, accommodating up to 20 microliters. The device boasts connectivity through a LAN port and four USB ports, and results are printed using a thermal recorder. Designed for reliability, this analyzer operates efficiently under diverse environmental conditions, making it suitable for laboratories demanding high-precision, automated blood analysis. It delivers a throughput of 30 to 70 samples per hour, minimizing carryover and ensuring reliable results. The analyzer uses cyanide-free reagents for hemoglobin testing, enhancing safety, and requires a minimal sample volume of 9 microliters for whole blood analysis. Capable of measuring multiple parameters such as WBC, RBC, HGB, HCT, and PLT, the device also provides three histograms for detailed analysis. Its lightweight design makes it space-efficient and ideal for various clinical settings.

Most popular related searches

Automated Blood Analyzer AM-BAA10 offers dimensions of 410 mm x 300 mm x 400 mm, designed for precise and efficient blood analysis. It operates in whole blood mode with a sample volume of less than or equal to 9 microliters and prediluted mode with less than or equal to 20 microliters. Its connectivity includes a LAN port and four USB ports, with printouts generated via thermal recorder. This analyzer is designed for reliability, operating efficiently under varying environmental conditions. Additionally, it is ideal for laboratories requiring high-precision, automated blood analysis.

  • Sample Volume (Whole Blood Mode) : Less than or equal to 9 uL
  • Sample Volume (Prediluted Mode) : Less than or equal to 20 pi-
  • WBC (107L) Performance : Linearity Range: 0 to 100 Precision (CV%): Less than or equal to 3.5% (4.0 to 6.9) Carryover: Less than or equal to 0.5%
  • RBC (10'VL) Performance : Linearity Range: 0 to 8.00 Precision (CV%): Less than or equal to 2.0% (3.50 to 6.5) Carryover: Less than or equal to 0.5%
  • HGB (g/L) Performance : Linearity Range: 0 to 280 Precision (CV%): Less than or equal to 2.0% (100 to 180) Carryover: Less than or equal to 0.5%
  • MCV (fL) Performance : Precision (CV%): Less than or equal to 1.0% (70.0 to 110.0) Carryover: <0.5%
  • PLT (10"/L) Performance : Linearity Range: 0 to 1000 Precision (CV%): Less than or equal to 5.0% (100 to 500) Carryover: Less than or equal to 1.0%
  • Hemoglobin Test : Cyanide-Free Reagent
  • Analyzed Parameters : 20 Parameters: WBC, Lymph#. Mid#, Gran#, Lymph%, Mid%, Gran%, RBC, HGB, HCT, MCV, MCH. MCHC, RDW-CV, RDW-SD, PLT, MPV, PDW, PCT, P-LCR
  • Throughput : 30 Samples Per Hour
  • Display : 8.4-inch TFT Touch Screen
  • Data Storage Capacity : Up to 50,000 Results
  • Communication Ports : LAN Port (HL7 Protocol, Bi-directional LIS)
  • USB Ports : 4 USB ports
  • Printout : Thermal Recorder, 50 mm Wide Paper, Multiple Printout Formats
  • Operating Temperature : 15*Cto30eC
  • Operating Humidity : 10% to 90%
  • Operating Air Pressure : 70kPatol06kPa
  • Power Supply : (100V to 240V) ±10%, Less than or equal to 180VA, (50Hz/60Hz) ±lHz
  • Dimensions (D x W H) : 410 mm x 300 mm x 400 mm
  • Minimal Sample Volume
  • High Accuracy and Precision
  • Low Carryover
  • Efficient Processing
  • User-Friendly interface
  • Large Data Storage
  • Multiple Connectivity Options
  • Integrated Printing
  • Stable Performance in Various Conditions

Our Automated Blood Analyzer AM-BAA10 is ideal for use In hospitals, diagnostic laboratories, research institutes, and blood banks, providing high-precision, automated blood analysis with minimal sample volume.