Intended Use: These oxygen sensors, with their excellent stability, performance and linearity are designed to be used to monitor the partial pressure of oxygen in breathing gas mixtures for anaesthesia, ventilators, medical oxygen concentrators, incubators, and general oxygen monitors.
Most popular related searches
medical oxygen sensor
medical oxygenator
medical oxygen
medical oxygen concentrator
medical sensor
breath gas
breathing gas
anesthesia ventilator
medical oxygen generating
medical oxygen generator
- Sensor Technology: Electrochemical
- Signal Output1: 9 - 14 mV
- Measuring Range: 0 - 100 Percent
- Response Time T90: 12 Seconds
- Accuracy Full Scale2: ± 1% of Signal
- Repeatability F.S.: ± 0.5%
- Temperature: Compensated
- Operating Temp: 0 to 45º C
- Storage Temp: 0 to 45º C
- Cross Sensitivity: ISO 80601-2-55
- Housing Material: White ABS
- Humidity: 0 - 100% RH (Non - Condensing)
- Expected Life3: < 60 Months
- Warranty4: 16 Months
- Signal output is measured in air at 25 º C, sea level.
- Full Scale accuacy is calculated with constant pressure, temperature and proper calibration (80% O2 value on full scale range). Drastic temperature change can result in a maximum error of ± 10%.
- Expected life is calculated when O2 < 20.9% @ 25 º C, sea level.
- Southland Sensing Ltd. warrants the sensors for the period noted above to be free from defects in materials and workmanship. Southland Sensing Ltd. will not be held liable for sensors damaged due to customer neglect.


- Vyaire / Carefusion Avea
- Vyaire / Carefusion Sipap
- GE Ivent Ventilator
- Versamed eVent Ventilators
- Siemens 300 / 900 Series Ventilator
- Sensor Output: 9 – 14mV
- Sensor Life: < 60 Months @ 20.9% O2
- Electrical Connection: 3-Pin Molex Header