RTI Group AB
  1. Companies
  2. RTI Group AB
  3. Products
  4. RTI - Fluoroscopy Dose Probe

RTIFluoroscopy Dose Probe

SHARE

The RTI Dose Probe is an external dose probe designed to perform very low dose rate measurements, for instance, on image intensifiers.

Most popular related searches

For low dose rate measurements

The probe is very small to avoid or minimize interference with AEC (Automatic Exposure Control) on X-ray equipment. This also enables it to fit into the table bucky.

Since it is a solid-state detector, no corrections for pressure or temperature are needed. Neither is polarizing voltage.

Dose probe for Fluoroscopy

The RTI Dose Probe has a fast response which makes it ideal for pulsed fluoroscopy. It can detect individual pulses, determine pulse rate, and show waveform at dose rate.

RTI Dose Probe specifications with Piranha

  • Size: 20 x 45 x 7.4 mm, 0.79" x 1.8" x 0.29"
  • Weight: 85 g (3 oz)
  • Cable length: 2.0 m (6.6 ft)
  • Backscatter protected: Yes
  • Dose: 100 pGy – 1.5 kGy. 12 nR – 170 kR (± 5%)
  • Dose rate: 4 nGy/s – 150 mGy/s (±5 % or ±10 nGy/s).
    • 0.46 μR/s – 16 R/s (±5 % or ±1 μR/s).
    • 26 μR/min – 1000 R/min (±5 % or ±6 μR/min).
  • Time: 0.1 ms – 34000 s (±1 % or ±0.5 ms)
  • Pulses: 1 – 65535 pulses (±1 pulse)
  • Dose per pulse: 1 nGy/pulse – 0.3 Gy/pulse
  • Pulse rate/frequency: 0.5 – 100 Hz
  • Pulse width: 4 ms - 2000 s

RTI Dose Probe specifications with Cobia FLEX & SENSE

  • Size: 20 x 45 x 7.4 mm, 0.79" x 1.8" x 0.29"
  • Weight: 85 g (3 oz)
  • Cable length: 2.0 m (6.6 ft)
  • Backscatter protected: Yes
  • Dose: 6 nGy - 2 kGy. 660 nR - 250 kR (± 5%)
  • Dose rate: 220 nGy/s - 220 mGy/s. 1.5 mR/min - 1.5 kR/min (± 5%)
  • Time: 0.33 ms – 9999 s
  • Pulses: 3 – 9999 pulses
  • Dose per pulse: 1 nGy/pulse - 3 kGy/pulse. 114 nR/pulse - 342 kR/pulse
  • Pulse rate/frequency: 1/6 - 260 Hz