Signal Group UK Ltd
Signal - Model 250 -Portable Cooler / Dryer
The Signal Group 250 Portable cooler dryer removes water vapour from stack and exhaust emission gases, before analysis by ambient analysers.
Most popular related searches
process emissions monitoring
greenhouse gas monitoring
flue gas monitoring
carbon capture and storage
emissions monitoring
combustion emissions monitoring
agriculture emissions monitoring
gas turbine exhaust
gas emissions monitoring
continuous emissions monitoring system
The 250 features a semiconductor Peltier cooling system with an aluminium cooling block. Fitted into the block is then a high efficiency heat exchanger made of stainless steel. The condensed water is then removed using a continuous peristaltic pump.
The status of the cooler is indicated by an LED system with high temperature alarm, a water alarm and an indicator that the set temperature is being maintained.
- Applications: Agricultural emissions monitoring, Automotive, Biogas, Biomass combustion monitoring, Boiler and Combustion Performance, Boiler/stove emissions compliance, Carbon capture and storage (CCS), Catalyst Research, Combustion emissions monitoring, Combustion research, Continuous emissions monitoring (CEM), Controlled atmosphere storage, Direct Reduction Iron (DRI) process emission monitoring, Engine Exhaust, Ethylene production, Fermentation monitoring, Flue gas emissions monitoring, Fuel cell analysis, Gas purity, Gas turbine exhaust emissions, Greenhouse gas (GHG) measurement, Landfill, Medical gas purity, Mining, Modified atmosphere packaging (MAP), Process control, Process Emission Monitoring, Safety monitoring, Specialist Research, Stack testing, Tobacco Testing, Toxic Gas Measurement, Vehicle emissions
CONFIGURATION: Single channel
INLET: Maximum inlet temperature: 210oC
- OUTLET:
Sample: dewpoint 5oC ± 0.5oC - Hot air: exhausted to ambient
- Condensate: pumped to drain
- FLOW RATE: 3 l/min (maximum)
PLUMBING:
- Gas sample(s): 1/4” tube fittings
Condensate drain: 1/4” stainless steel - POWER: Voltage rating: 115 Vac or 230 Vac. Consumption: 180VA maximum
- ENVIRONMENT: Ambient temperature: + 5oC to 35oC. Max Humidity: 95% RH non-condensing
FRONT PANEL:
- Mains power on/off switch
- Each channel has the following LED indicators:
- Status (green)
- High temperature alarm (red)
- Low temperature alarm (red)
- Water alarm (red)
- Power (green)
ALARMS:
- High temperature.
- Low temperature
- Water condensate present
DIMENSIONS: 630mm x 500mm x 400mm