The well known analysers of the product line SUPER G were developed for clinical laboratory. With the SUPER GL speedy an analyser was developed which uses the modern technology of the sensors together with other new technical items. So the SUPER GL speedy has besides a innovative sample rotor for 60 samples a new touch screen for easy operating the device in menu choosing and together with both news also the well known biosensor.

Most popular related searches

As with all analysers using the enzymatic-amperometric measuring principle, pre-analysis is quite simple. Only a few µl of capillary blood are needed, taken from the finger berry using an end-to-end or open-end-capillary. These samples are diluted using a defined amount of solution in a sample cup and hemolysed, and can then be measured immediately. Approximately one minute after taking the blood, you have the result.

All devices provide the technical standard necessary for modern analysers, such as a variety of EDP links, barcode reader, storage function for the samples and controls etc.

We realised the IFCC - recommendation for displaying the glucose values plasma referenced at the SUPER GL easy+.

Performance Data

Measuring Time
Single Measurement Glucose: ca. 40 seconds
Single Measurement Glucose / Lactate / Hemoglobin: ca. 75 seconds

Measuring Range
Glucose: 0.5 mmol/l to 50 mmol/l or 9 mg/dl to 900.8 mg/dl
Lactate: 0.5 mmol/l to 30 mmol/l or 4.5 mg/dl to 270 mg/dl
Hemoglobin: 1.86 mmol/l to 15.52 mmol/l or 3 g/dl to 25 g/dl

Sample Volume
10 µl sample diluted with 500 µl hemolysate solution

Precision (20 samples) - Variation Coefficient Indication
Glucose (12 mmol/l or 216 mg/dl): < 2.0 %
Lactate (10 mmol/l or 90 mg/dl): < 2.5 %
Hemoglobin (7.6 mmol/l or 12 g/dl): < 2.0 %

Sensors
Operating Time Glucose Sensor: max. 6 months / max. 10.000 samples
Operating Time Glucose and Lactate Sensors: max. 3 months / max. 10.000 samples

Technical Specifications
Working Temperatur: + 15 °C bis + 35 °C
Supply Voltage: 12 V DC
Power Consumption: max. 45 W
Classification according MPG: In-vitro-Diagnosticum (guidline 98/79/EC)

Interfaces
Printer: serial, RS 232
EDP: serial, RS 232
Barcode Reader: external, PS/2

Dimensions / Weight
Width: 220 mm
Hight: 170 mm
Depth: 185 mm
Weight: 1.7 kg