Biosen - Model C-Line - Glucose and Lactate Analyzers for Athletes
The Biosen series is the leading brand name in lactate and glucose threshold measurement for top class athletic sports institutes and professional sports teams.
Biosen analyzers can be manufactured as single channel lactate systems or as dual channel systems that can measure glucose at the same time.
Biosen uses a special chip sensor technology to deliver fast measurements with a high degree of accuracy at a low cost per test. Due to the long life of the sensor chip the system requires little maintenance, features such as the touch screen and large memory only add to the Biosen analyzer’s user-friendliness.
An integrated barcode reader and printer are also available with most configurations, allowing entire teams of athletes to have multiple step-tests during training.
Biosen can test blood, plasma or serum to provide lactate and glucose values with excellent precision (less than 2% CV) over a wide measurement range.
Easy glucose and lactate measurements in 3 steps
- Collect capillary blood sample.
- Place capillary into cup and mix.
- Result in 20-45 seconds
- Easy to use
- Results in 20-45 seconds
- Up to 120 results an hour
- Imprecision < 1.5% CV
- Size: 280 x 280 x 95 mm
- Multi-lingual touch screen display with step by step instructions
- Just 20 μl blood, plasma or serum sample required when used with EKF pre-filled `Safe-Lock` reaction cups (1,000 μl)
- Up to 120 results an hour
- Results in 20-45 seconds
- Automatically detects presence of reaction cups
- Chip sensors easily exchangeable, no electrode cleaning required
- Enzymatic-amperometric method using chip-sensor technology
- Measuring range: Lactate 0.5–40 mmol/L (5–360 mg/dL), Glucose 0.5–50 mmol/L (9–900 mg/dL)
- Imprecision: CV ≤1.5 % (12 mmol/L)
- Pre-filled ‘Safe-Lock’ reaction cups eliminate manual dilution
- Innovative needle and exchanger design eliminates cross contamination
- Three calibration modes
- Dual channel option: Glucose and/or lactate results from one sample
- Long sensor life: Glucose 60 days / 7,500 tests; Lactate 50 days / 6,000 tests
- Sample stable for up to five days (refrigeration recommended)
- STAT position
- PC, network and printer connectivity via RS232C interfaces
- Stores up to 1,000 results
- Integrated barcode reader (optional)