Heska Corporation
  1. Companies
  2. Heska Corporation
  3. Products
  4. Cuattro - Model DR HD - High Definition ...

CuattroModel DR HD - High Definition Digital Radiography System

SHARE

A smarter direction in small animal digital radiography.

Most popular related searches
  • Exceptional Images

    Delivering detail, resolution, and contrast image quality that is visibly better than Standard Definition systems.

  • Industry Leading Quality

    From the leading designers in veterinary digital radiography for more than a decade.

  • An Incredible Improvement Over Film or CR

    Avoid the high cost, retakes, over-exposure, processing failures and storage related to outdated technology.

  • Easier Workflow and Better Images

    Shorten exam times while getting superior diagnostic images.

  • Increased Profitability

    Do more exams, order more views, consult with superior images, and diagnose patients better.

  • Cloud Based Storage with Cloud DR

    Save studies to the cloud for immediate access and ability to share with anyone, anywhere.

CloudDR High Definition Digital Radiography HD Minimum Standards
  • Direct Deposition Cesium
  • 100 micron pixel size
  • HD pixel density. 18 Million (17×17)
  • HD pixel density. 15 Million (14×17)
  • Full Size Image Area (14×17 – 17×17)
  • Resolution of at least 5 lp/mm
  • 16 Bit A/D 16 Bit Grayscale Native
  • 16 Bit Multi-Frequency Tuning

CloudDR™ HD Added Benefits

  • Automatic Exposure Detection
  • Cassette Sized Form Factor
  • Windows Embedded – 64 bit OS
  • Multi-Touch User Interface
  • Cloud Based PACS Connectivity
  • Comprehensive Warranty w/ Cloud Archival
  • Software Upgrades and Updates
  • 24/7 Live Technical Support

Sensor

  • Panel – Single substrate amorphous silicon active TFT/diode array
  • Scintillator – Direct deposition CsI:Tl
  • Pixel Matrix – 3556 × 4320
  • Pixel Pitch – 100 μm

Electronics

  • Amplifiers – Low noise ASICs with user selectable gains
  • ADC – 16 bit
  • Image Transfer Time Wired – 500 ms; Wireless: 3000 ms
  • On Board Memory – 1 GB DDR3, 4 GB SDHC card

Mechanical

  • Size – ISO 4090 for 35 cm × 43 cm (14” × 17”) cassette size
  • Active Area – True 355 mm × 432 mm
  • External Dimensions – 384 mm (w) × 460 mm (l) × 15 mm (h)
  • Weight – 3.8 kg (8.4 lbs)
  • Housing Carbon – fiber front & back

Communications

  • Status Display – OLED display with WiFi, battery, and sensor indicators
  • Wireless Data I/F – 802.11n WiFi standard
  • Wired Data I/F – GigE via optional power & communication tether
  • X-ray I/F – Integrated X-ray trigger control Automatic Exposure Detection

Imaging Performance

  • Limiting Resolution – 5 cy/mm
  • Typical MTF – 70% (1 cy/mm), 40% (2 cy/mm), 15% (4 cy/mm) for RQA5
  • Typical DQE – 75% (0 cy/mm), 60% (1 cy/mm), 40% (3 cy/mm) for RQA5

Environmental

  • Temperature: 10 – 35 °C operating
  • Humidity: 30 – 70 % RH operating (non-condensing)

Power

  • Battery – Rechargeable battery, 53.3 Wh
  • Battery Charger – External two bay charger 100-240 V AC 50/60 Hz
  • Interface & Power Unit – Optional XRpad IPU with external power supply 100-240 V AC, GigE, and X-ray I/F

14×17” Wireless HD Detector Specifications: Sensor

  • Panel – Single substrate amorphous silicon active TFT/diode array
  • Scintillator – Direct deposition CsI:Tl
  • Pixel Matrix – 3556 × 4320
  • Pixel Pitch – 100 μm

Electronics

  • Amplifiers – Low noise ASICs with user selectable gains
  • ADC – 16 bit
  • Image Transfer Time Wired – 500 ms; Wireless: 3000 ms
  • On Board Memory – 1 GB DDR3, 4 GB SDHC card

Mechanical

  • Size – ISO 4090 for 35 cm × 43 cm (14” × 17”) cassette size
  • Active Area – True 355 mm × 432 mm
  • External Dimensions – 384 mm (w) × 460 mm (l) × 15 mm (h)
  • Weight – 3.8 kg (8.4 lbs)
  • Housing Carbon – fiber front & back

Communications

  • Status Display – OLED display with WiFi, battery, and sensor indicators
  • Wireless Data I/F – 802.11n WiFi standard
  • Wired Data I/F – GigE via optional power & communication tether
  • X-ray I/F – Integrated X-ray trigger control Automatic Exposure Detection

Imaging Performance

  • Limiting Resolution – 5 cy/mm
  • Typical MTF – 70% (1 cy/mm), 40% (2 cy/mm), 15% (4 cy/mm) for RQA5
  • Typical DQE – 75% (0 cy/mm), 60% (1 cy/mm), 40% (3 cy/mm) for RQA5

Environmental

  • Temperature – 10 – 35 °C operating
  • Humidity – 30 – 70 % RH operating (non-condensing)

Power

  • Battery – Rechargeable battery, 53.3 Wh
  • Battery Charger – External two bay charger 100-240 V AC 50/60 Hz
  • Interface & Power Unit – Optional XRpad IPU with external power supply  100-240 V AC, GigE, and X-ray I/F