Arrowhead Medical Device Technologies, LLC

Arrow-Lok - Implant Technology
Design Rationale: In planning and research prior to design of the ARROW-LOK™ implant, the surgeon design team identified performance objectives for the new implant that would offer clinical advantages over current implant alternatives.
Most popular related searches
implant technology
implantable medical device
implant
patient anatomy
implant engineering
implant insertion
implant placement
bone removal
medical device
anatomy
These design imperatives included:
Stable fixation
- Keep completely intramedullary with no parts of the implant exposed outside the skin.
- Maintain the initial compression applied at insertion.
- Resist the rotational and pull-out forces affecting the lesser toes.
Simple implant design
- Avoid introducing potential stress-risers and the need to assemble intraoperatively by developing a one-piece design.
- Provide anatomic sizing to fit a broad range of patient anatomy.
- Fit the patient’s natural toe anatomy by providing a neutral and 10° plantar angle.
- Minimize bone removal with a low-profile design.
- Maintain uniform strength characteristics throughout the implant.
- Ensure the product material has a proven long-term biocompatibility so it can be safely implanted permanently. Will not require removal like the K-wire.
Efficient technique
- Create an operative technique similar to that for K-wire fixation with little additional complexity, such as drilling, tapping, or bulky instrumentation.
- Remove little or no bone when preparing the bone for implant insertion while compacting cancellous bone to enhance stable fixation.
- Enable variations in implant placement with respect to the fusion site without any impact on implant effectiveness.
- Allow enough versatility within the interaction between instrument and implant such that the technique may adapt to surgeon preference and patient anatomy.
Proven material
- Must be biocompatible, clinically proven, extensively studied, and commonly used for permanently implanted medical devices.
- Bending and fatigue characteristics able to endure the forces exerted on the lesser toes without having to add unnecessary material volume.
- Capable of being efficiently manufactured to ensure reproducible results.
- Does not require special handling or storage either preoperatively or intraoperatively.
Simple instrumentation
- Instrumentation resembles those with which surgeons are familiar.
- Emphasize simplicity to ensure reproducible results.