VITALMAX - Model 4000SL - Medical Patient Monitors
The VITALMAX 4000SL is designed with an optimal ability against electrosurgical interference. It offers a simple and adaptable interface. Patient information along with vital sign settings can be quickly modified to meet the needs of a patient’s changing condition. The VITALMAX 4000 offers a high resolution 12″ color LCD touch screen for making quick screen adjustments, set default settings, alarm limits, and manage up to 72 hours of detailed patient data. It boasts a very short recovering time after defibrillation, especially suitable for use in an emergency or operating room.
- 12-inch touch screen
- Optional EtCO2, Dual IBP
- 2x IBP & CO
- 2x IBP & EtCO2
- EtCO2 Only
- CO Only
- Masimo SpO2
- 5-lead EKG (with additional 3-lead operation)
- SunTech Non-Invasive Blood Pressure (NIBP)
- Temperature Monitoring
- Integrated Thermal Printer
From the patient room to high acuity surgeries, the VITALMAX 4000 monitors are designed to fit in and move to any patient care areas easily. It offers standard measurements of non-invasive blood pressure, ECG, SpO2, Temperature, and Respiration rate. End-tidal CO2 can be added on-site by simply attaching our plug in modules. Dual Channel Invasive Blood Pressure (IBP) is also optional.
The 12’’ color TFT LCD screen has adjustable brightness, a wide viewing range, multiple waveforms can be displayed on the same screen, and has volume control. The streamlined outline design has a handle, along with a multi-language interface selection.
It contains a unique thoracic impedance method. Advanced digital technology accurately measures SpO2 in low perfusion situation.
The user-friendly design offers a dialogue interface with pop-up menus, dialogue boxes, rotary mouse, low-power consumption with standby mode, rechargeable high-energy built-in battery, blood pressure measurements that are fast and accurate.
It has a drug calculation, ST-segment detection display format that includes OxyCRG diagram, 5-lead ECG waveform, and large digits.
The sound and light integrated alarm parameters can be adjusted easily.
It uses a WAN communication function to network with a central monitoring system so it is efficiently used to make long-distance monitoring, diagnosing, and patient maintenance achievable, and a software upgrade is possible.
Physical Weight & Dimensions
- Weight: 5kg
- Height: ~11 inches
- Depth: ~6-1/5 inches
- Width: ~12 3/5 inches
Application
- Neonatal, pediatric and adult patients
Operation Environment
- Power Source - External AC power or internal battery
- AC Power - 100-240VAC, 50/60Hz, <150VA
- Battery - Rechargeable Lead-Acid; Operating time under the normal condition (1 hour); Operating time after the first alarm of low battery (10 min.)
Temperature
- Working - 5°-40° C
- Storage - -20°-65° C
Performance Specifications
- Display - 12.1" color TFT, Touch Screen
- Resolution - 800 × 600 pixels
- Trace - 4, 6 or 8 waveforms ECG (I, II, III, aVR, aVL, aVF, V1-V6), PLETH, RESP, IBPx2, ETCO2
- Indicator - Alarm indicator; Power indicator QRS beep and alarm sound
- Trend Time - 72 hours
- Recorder - Built-in, thermal array, 2 channels
- Record - Width 48mm
- Recorder - Paper 50mm
- Record - Speed 25mm/s, 50mm/s
ECG
- Input - 3-lead ECG cable and standard AAMI line for connection
- Input - (Optional) 5-lead ECG cable and standard AAMI line for connection
- Lead Choice - I, II, III, aVR, aVF, aVL, V
- Gain Choice - x0.5, x1.0, x2.0
- Frequency Characteristic - 0.05-40 Hz
- ECG Waveforms - 7 channels
- Penetration Voltage - 4000VAC 50/60Hz
- Sweep Speed - 12.5, 25 and 50 mm/s
- HR Display Range - 30-300bpm
- Accuracy - ±1% or ±1bpm, whichever is greater
- Alarm Limit Range Setting - Upper limit 80-400bpm; Lower limit 20-150bpm
RESP
- Measure Method - RA-LL impedance
- Range - 0-120 rpm
- Accuracy - + 3 rpm
- Alarm Upper-lower Limit Setting - Upper limit 6-120 rpm, Lower limit 3-120 rpm
- Sweep Speed - 12.5 and 25 mm/s
NIBP (Suntech)
- Measuring Technology - Automatic oscillating measurement
- Cuff Inflating - <30s-0-300 mmHg, standard adult cuff
- Measuring Period - AVE<40s
- Mode - Manual, Auto, STAT
- Measuring Interval in AUTO Mode - 2 min-4 hrs
- Pulse Rate Range - 30-250 (bpm)
- Measuring Range - Adult/Pediatric Mode, SYS 40-250 (mmHg), DIA 15-200 (mmHg); Neonatal Mode, SYS 40-135 (mmHg), DIA 15-100 (mmHg)
- Resolution - 1mmHg
- Accuracy - Pressure, Maximum Mean error ±5mmHg, Max. Standard deviation 8mmHg
- Overpressure Protection - Adult Mode 280(mmHg), Neonatal Mode 150 (mmHg)
- Alarm Limit Setting - SYS 50-240 mmHg, DIA 15-180 mmHg
TEMP
- Range - 25°-50° C
- Accuracy - + 0.2 °C 25.0°-34.9° C; + 0.1° C 35.0°-39.9° C; + 0.2° C 40.0°-44.9° C; 0.3° C 45.0°-50.0° C
- Display Resolution - 0.1° C
- Alarm Upper-lower Limit Setting - Upper limit 0-50° C, Lower limit 0-50° C
- Channel - 2 channels
- Alarm Limit Setting - 10-50° C
Masimo SET SpO2
- SpO2 Accuracy (non-motion):
- Adult Pediatric: 70~100 %: ±2 %, 0~69%: unspecified
- Neonate: 70~100 %: ± 3 %,0~69 %: unspecified
- SpO2 Accuracy (motion):
- Adult Pediatric: 70~100 %: ±3 %, 0~69%: unspecified
- Neonate: 70~100 %: ± 3 %, 0~69 % : unspecified
- Low Perfusion Performance: >0.02 % Pulse Amplitude and % Transmission >5
- SpO2: ±2 %
- PR: ±3 bpm
- Modes:
- Averaging mode: 2,4,8,10,12,14 and 16 s
- Sensitivity: Normal, APOD and Maximum
- PR Accuracy (non-motion):
- Neonate: 70~100 %: ± 3 %, 0~69 % : unspecified
- Adult Pediatric Neonate: 25~240 bpm: ±3 bpm
- PR Accuracy (motion):
- Adult Pediatric Neonate: 25~240 bpm: ±5 bpm
- Measuring Range:
- SpO2: 1~100 %
- PR: 25~240 bpm
- Perfusion: 0.02~20 %
IBP (Option)
- Measurement Range - -10-300mmHg
- Channel - 2 channels
- Pressure Transducer Sensitivity, 5μV/V/mmHg
- Impedance Range - 300-3000Ω
- Transducer Sites - ART, PA, CVP, RAP, LAP, ICP
- Resolution - 1mmHg
- Accuracy - ±1mmHg or ±2%, whichever is greater
- Alarm range - -10-300mmHg
EtCO2 (Option)
- EtCO2 Option allows compliance with
- Mode of Sampling – Sidestream
- Principle of Operation -Non-dispersive infrared (NDIR) single beam optics, dual wavelength, no moving parts
- CO2 Measurement Range - 0 to 150 mmHg (0 to 19.7%, 0 to 20 kPa)
- CO2 Calculation Method - BTPS (Body Temperature Pressure Saturated)
- CO2 Resolution - 0.1mmHg(0-69mmHg); 0.25mmHg(70-150mmHg)
- CO2 Accuracy - 0-40 mmHg ± 2 mmHg; 41-70 mmHg ± 5% of reading; 71-100 mmHg ± 8% of reading; 101-150 mmHg ± 10% of reading; Above 80 breath per minute ± 12% of reading
- Sampling Rate - 100Hz
- Respiration Rate - 2-150 bpm
- Respiration Rate Accuracy - ±1 breath
- Response Time - <3 seconds - includes transport time and rise time
- Inspired CO2 Measurement Range - 3-50 mmHg