AbluminusModel DES + - Sirolimus Eluting Coronary Stent System

SHARE

ABLUMINUS DES+ designed by propriety ENVISOLUTION technology in a way that address the unmet clinical needs specially in Diabetes Mellitus and Acute Myocardial infarction patients with Coronary Artery Disease. ABLUMINUS DES+ Receives on-label indication for Diabetes Mellitus and Acute Myocardial infarction.

Most popular related searches

Envisolution Technology

ABLUMINAL COATING
Facilitate mono directional drug release and less systemic exposure of drug leading to faster re-endothelistation

FUSION COATING
Coating on the stent and exposed parts of the balloon facilitate homogenous drug delivery which address diffused proliferative disease and focal restenosis

EDGE COATING
The additional 0.5mm coating on the edge of the balloon which addresses the edge restenosis

BIODEGRADABLE FILM
The formation of circular film with biodegradable polymer facilitate maximum drug delivery in blood wet conditions

Drug Polymer
Drug: Sirolimus
Drug Dose: 0.70µg/mm²
Drug Carrier: Customized biodegradable polymer matrix

Stent
Stent Material: Cobalt Chromium Alloy L605
Strut Thickness: 73µm
Strut Width: 80µm (hinge) – 120µm (strut)
Average Recoil: Less than 5%
Average Foreshortening: Less than 1% (Zero foreshortening design)

Delivery System
Delivery System: RX stent delivery system
Nominal Pressure: 8 bar
Rated Burst Pressure: 14 bar
Guide Wire Compatibility (max.): 0.014″
Guiding Catheter Compatibility: 5F
Stent Diameter (mm): 2.25, 2.50, 2.75, 3.00, 3.50, 4.00
Stent Length(mm): 8, 12, 16, 20, 24, 28, 32, 36, 40
Crossing Profile: 0.038″
Entry Profile: 0.016″